The primary method for sterilizing instruments in the medical office is ____.

Learn the purpose and limitations of autoclaves, types of cycles, and procedures for safe and effective autoclaving.

Purpose

Autoclaving, sometimes called steam sterilization, is the use of pressurized steam to kill infectious agents and denature proteins. This kind of "wet heat" is considered the most dependable method of sterilizing laboratory equipment and decontaminating biohazardous waste.

Other decontamination methods—dry heat, ultraviolet or ionizing radiation, and liquid, gas, or vapor disinfection—are not a proper substitute for autoclaving or incineration before disposing of biohazardous material. Autoclaves do not remove chemical contamination.

Operating procedures

When used properly, autoclaves are safe and highly effective. Autoclaves use saturated steam under pressure of approximately 15 pounds per square inch to achieve a chamber temperature of at least 250°F (121°C) for a prescribed time—usually 30–60 minutes.

In addition to proper temperature and time, prevention of entrapment of air is critical to achieving sterility. Material to be sterilized must come in contact with steam and heat.

Using an autoclave requires caution and adherence to strict regulatory and operational requirements. For operating procedures, read:

Autoclave cycles

There are 2 basic autoclave cycles:

  • Gravity or "fast exhaust"
  • Liquid or "slow exhaust"

Both cycles and the materials appropriate for each cycle are described below.

Cycle

Materials

Description

Gravity or "fast exhaust"

Dry goods, glassware, etc.This cycle charges the chamber with steam and holds it at a set pressure and temperature for a set period of time. At the end of the cycle, a valve opens and the chamber rapidly returns to atmospheric pressure. Drying time may also be added to the cycle.

Liquid or "slow exhaust"

LiquidsThis cycle prevents sterilized liquids from boiling. Steam is exhausted slowly at the end of the cycle, allowing the liquids (which will be super-heated) to cool.

Sterility monitoring

Chemical indicator (e.g., autoclave tape) must be used with each load placed in the autoclave. However, use of autoclave tape alone is not an adequate monitor of efficacy. Autoclave sterility monitoring must be conducted at least monthly using appropriate biological indicators (Bacillus stearothermophilus spore strips) placed at locations throughout the autoclave.

The spores, which can survive 250°F for 5 minutes but are killed at 250°F in 13 minutes, are more resistant to heat than most, thereby providing an adequate safety margin when validating decontamination procedures. Each type of container employed should be spore tested because efficacy varies with the load, fluid volume, etc.

Help UCSD get the lead out!

Some brands of autoclave indicator tape may contain lead. Learn more about possible high levels of lead in autoclave tape, how to properly dispose of it, and lead-free alternatives.

Notice: Disposal of hazardous waste using sinks, intentional evaporation, or as regular trash is against the law. Campus laboratories must abide by strict state and federal waste disposal requirements. You may be held liable for violations of applicable laws.

The safe provision of oral health care services is a top priority for both patients and providers. Ensuring that infection control protocols are precisely followed is the responsibility of all dental team members. While infection ­con­­trol protocols encompass a wide array of policies and procedures, the sterilization of dental instruments is a critical component of ensuring safety in the dental operatory. In dentistry, there are three ways to sterilize instruments: steam under pressure, dry heat, and chemical vapor. In this informative Q&A, infection control expert Chris H. Miller, PhD, shares his insight on the dry heat sterilization method.

How does dry heat sterilization work?

Dry heat sterilization causes proteins to lose their distinctive shapes as microbes dehydrate. Free oxidative radicals are created as microbial proteins coagulate and microbial death ensues. The amount of water in and around microbes influences the temperature at which proteins are destroyed. The lower the water content, the higher the temperature needed for destruction.1 This is why steam sterilizers, which use moist heat, operate at lower temperatures (about 121° C to 134° C) compared to dry heat sterilizers (about 160° C to 190° C).2

The primary method for sterilizing instruments in the medical office is ____.

It’s important to ensure that the packaging materials and the instruments and other items being processed can withstand the high temperatures of dry heat sterilization. As with any sterilizer, successful sterilization occurs only after the prescribed exposure time has been achieved at the indicated sterilizing temperature. So “warm-up” time (time required for the chamber to reach sterilizing temperature) is not included in the sterilizing exposure time. Also, interrupting the exposure time—such as opening the door of a dry heat sterilizer mid-cycle (if allowed) to add a forgotten item—circumvents the sterilization process. Only equipment that has been cleared by the United States Food and Drug Administration (FDA) as a sterilizer should be used to process dental instruments, and the manufacturer’s directions need to be followed precisely to assure successful sterilization.3

What are the two types of dry heat sterilizers and how are they different?

Static-air and forced-air are the two types of dry heat sterilization, and they are based on how the hot air circulates in the sterilizing chamber.

The static-air (gravity convection) type depends on natural air convection. As the air is heated by coils on the bottom or sides of the chamber, it rises and the cooler air falls to create circulation. The sterilizing time usually ranges from 60 minutes to 120 minutes (after the warm-up time is complete), depending on the equipment used.

Forced-air (mechanical convection) uses a blower motor to mechanically circulate the chamber air, which decreases the heat-energy transfer time from the hot air to the items being processed. Thus, the sterilizing cycle is shorter than static-air models and may be as brief as 12 minutes for wrapped instruments, depending on the equipment used.

What are the benefits of dry heat sterilization?

A major benefit of dry heat sterilization is minimal or no instrument corrosion. Oxygen, moisture, and high temperature facilitate corrosion of susceptible metal items (eg, carbon steel). In dry heat sterilization, moisture is eliminated, so corrosion is minimized. This is why it’s especially important to dry the cleaned instruments prior to packaging and using dry heat sterilization. Another benefit is that instrument packages are dry at the end of the sterilization cycle and are ready to use after the appropriate cooling time. This feature of forced-air dry heat sterilization helps provide a relatively short instrument turnaround time, depending on the equipment used.

How does dry heat sterilization compare to steam sterilization?

There are a variety of FDA-cleared steam and dry heat sterilizers available. Steam sterilization (ie, autoclaving) is dependable and economical and is the most widely used sterilization method for items that are not sensitive to heat and moisture. Autoclaving can cause corrosion of nonstainless steel (eg, carbon steel) items such as burs and some cutting instruments. Also, the sterilized items are wet at the end of a steam sterilizing cycle.

Dry heat sterilization can be used on items that corrode in steam or are impenetrable to moist heat (eg, powders, petroleum products).3,4 The sterilized items are dry at the end of a dry heat sterilizing cycle. The static-air dry heat sterilizers have longer sterilizing cycles than steam sterilizers. Some forced-air dry heat sterilizers, however, have been reported to provide shorter instrument turnaround times than steam sterilizers.5

FDA-cleared packaging materials and chemical monitoring devices are available for both methods of sterilization. Steam and dry heat sterilizers can be biologically monitored with appropriate bacterial spores, including Bacillus atrophaeus for dry heat and Geobacillus stearothermophilus for steam. This step is important in ensuring the success of the sterilization process.

Dry heat sterilization is an effective method for sterilizing instruments in the dental setting, and may be particularly well suited for practices in which instrument corrosion is an issue.

Which method of sterilization is the most widely used in the medical office?

Of all the methods available for sterilization, moist heat in the form of saturated steam under pressure is the most widely used and the most dependable. Steam sterilization is nontoxic, inexpensive 826, rapidly microbicidal, sporicidal, and rapidly heats and penetrates fabrics (Table 6) 827.

What is the first step in sterilizing instruments?

Decontamination: All instruments and trays must be pre-cleaned using water mixed with either detergent or specialized cleaning solution to remove foreign materials before placing them into a mechanical cleaning machine to be washed and dried.

Which sterilization method is most often used in a medical office quizlet?

Although in medical offices steam under pressure is the most frequently used method of sterilization, the method depends on the nature of the material to be sterilized and the type of microorganisms to be destroyed.

Which methods are used for instrument sterilization?

Steam Sterilization..
Flash Sterilization..
Low-Temperature Sterilization Technologies..
Ethylene Oxide “Gas” Sterilization..
Hydrogen Peroxide Gas Plasma..
Peracetic Acid Sterilization..
Microbicidal Activity of Low-Temperature Sterilization Technologies..
Bioburden of Surgical Devices..